Brain components are thought to be generated by patches of cortex across which local field activity becomes spatially coherent.
The electrical field of this patch can usually be modeled accurately by an "equivalent current dipole"
(ECD), which is like a tiny, electrical version of a bar magnet. This means the electrical
field produced by the component would create a positive potential on one side of the ECD and a
negative potential on the other. Therefore, fitting a theoretical dipole to the scalp topography
should result in a good fit. Some components are better described by 2 dipoles (meaning two distinct
patches of cortex are synchronized), and so the one dipole fit shown in the images may have higher
residual variance than expected.
While brain activity exists at frequencies that surpass 200 Hz,
low frequencies are typically the only ones synchronous enough to show up in EEG.
Therefore, brain components tend to have diminishing power at higher frequencies. Additionally,
brain components tend to have repeated patterns at certain frequencies, leading to a peak in the
power spectrum. These peaks are often found between 5 and 30 Hz, with 10Hz (termed alpha) being the
most common.
As far as the component time series go, it's difficult to know what to expect unless the dataset
is epoched (i.e. repeated sections of the experiment have been preselected). If the data is epoched,
you can look for an event related potential (ERP). You can tell if a component comes from an epoched dataset
by reading the label above the ERP Image.The best way to see if there is an ERP is by looking at he ERP-Image
plot on the top right of the image. In the lower section of that plot, is an average of all trials.
Summary
Scalp topography often looks dipolar
Residual variance from dipole fit (marked RV on images) should be low.
Usually below 15% unless the component is better explained with two dipoles
Power spectrum decreases as frequency increased (1/f)
Power spectrum usually has peaks between 5 and 30 Hz, most often at 10 Hz
Epoched data will likely have a visible ERP
Examples
Eye Component
Eye components describe eye motion. Each retina (the part of the eye that registers incoming light)
creates an electric field which can be effectively modeled as an "equivalent current dipole"
(ECD). Typically, eye motion is split into two components: vertical movement and horizontal movement.
Other eye components can also be found, such as diagonal directions, but they are rare and depend on
the experiment. For all eye components, the power spectrum will vary due to experiments and people, but
generally most of the power will reside at frequencies below 5 Hz as people do not usually move their eye faster
than that.
Vertical movement has a scalp topography that can be modeled with two ECDs, on in each eye, that are
oriented up down and there for only look positive or only negative as EEG recordings rarely record from
the underside of the head. Due to nearness of the eyes and the strength of the ECDs, one ECD can usually
fit this type of component very well. The component activity should show clear spikes relatively frequently
due to eye-blinks. There can also be some drift due to looking up or down.
Horizontal eye movement has a scalp topography that can be modeled an ECD placed between the eyes
and oriented left/right with positive values on one side and negative on the other. The component
activity should have intervals of relative stability with occasional and very fast transitions to
different values. Such patterns are cause by visual scanning. The periods of stability are times when
visual focus is help somewhere and the fast transitions are saccades.
Summary
Scalp topographies suggest ECDs near eyes
Power concentrated at low frequencies (below 5 Hz)
Vertical eye movement components will contain blinks in the data
Horizontal eye movement components will look like step functions
Muscle Component
Muscle components describe the electrical fields generated by muscle activity, known
as electromyography (EMG).Their activations
are powerful relative to EEG but motor unit action potentials
(the underlying source of EMG) do not synchronize causing most of the power of EMG to be spread out
among higher frequencies. Nonetheless, these components can still look dipolar, although they will seem very shallow
as they are not localized within the brain. You can tell a shallow dipole by how concentrated its scalp topography is.
The more concentrated, the shallower. That isn't to say that all muscle components will be dipolar.
Summary
Power concentrated in higher frequencies (20 Hz and above)
Can still be dipolar, but will be located outside the skull
Heart Component
Heart components capture the electrical potentials generated by the heart. This measurement is known
as electrocardiography (ECG or EKG).
The pattern generated by the heart is very typical and is known as a
QRS complex . These should occur at about 1 HZ.
because of the distance of the heart, the scalp map will look like that of a very far dipole and so will
look almost like a linear gradient.
Summary
Clear QRS complex in the data at about 1 Hz
Near linear gradient scalp topography
No peaks in power spectrum
Line Noise
Line noise is contamination from the alternating current that is used to power nearly all lighting fixtures and electronics
these days. Depending on where you are, the frequency used can be either 50 Hz or 60 Hz. The datasets we use come from all over
and so you can expect to see some of both (but never together). While this noise is typically removed in the data cleaning
steps before ICA is applied, ICA may separate out line noise if it has enough channels and data to work with. Line noise
components are mostly evident from their sharply peaked power spectrum with the peak at either 50 or 60 Hz. These
components are not to be confused with other components that simply have some line noise contaminating them.
Note: To avoid issues with this type of noise, many datasets are passed through a notch filter centered at either 50 Hz
or 60 Hz. This will cause a sharp dip in the power spectrum at that frequency and is not indicative of any of these categories.
Summary
Strong peak in power spectrum at either 50Hz or 60Hz
Channel Noise
If a channel gets bumped a lot during a recording or if it has a poor contact, it will often generate
large artifacts that do not affect any other channels. ICA will often separate these out into their own component
which we call channel noise components. They can be recognized by their scalp topographies which put almost all the
weighting on a single channel.
These components can be difficulty to classify as often times they can seem very similar to muscle
components. The similarity resides mostly in the scalp topography. Telling them apart is primarily
accomplished by looking for typical high frequency power typical of muscle components. Channel noise
components, on the other hand, typically have a 1/f spectrum.
Summary
Very focal scalp topography
Large and/or consistent artifacts in the component activations.
Easily confused with muscle components, but PSD is different.
Other
Not all components are meaningful. More specifically: most components are not meaningful.
ICA assumes that there are as many independent components as there electrodes and that is almost never
the case. When a component does not converge onto a meaningful signal, it can either capture a mixture
of signals or some noise. In rough terms, anything that does not fit the above categories can be deemed
"other". Signs of components being in the "other" category would be anything that stands out as very unusual
or uncharacteristic of the other categories. Also, the high the "component number" (and therefore the lower
the explained variance in the data), the more likely that a component is in the "Other" category. Usually the
last half of components can be called "Other". For instance, even if there seems to be a 10 Hz peak in the
power spectrum, the component is not necessarily a brain component. If it has a very non-dipolar scalp
topography and there is very little
Summary
Anything that doesn't fit the above categories.
More likely the higher the IC number (as in IC 150 of 220 is very likely to be "Other"
Non-dipolar scalp maps
Spectrum can still have weak 10 Hz peak as brain signals are likely mixed with other signals